The Olympus DP30 is a high-sensitivity cooled monochrome digital CCD camera. With its easy-to-use software, high resolution (1360 x 1024 pixels), low noise images are acquired effortlessly. This camera is ideal for a broad range of low-light and fluorescence imaging applications requiring sensitivity
In addition, there is a 6 channel image merging function where fluorescence and DIC combinations can be merged into one multi-color image.
DP30拥有高灵敏度的黑白制冷CCD ,非常适合荧光等需要高灵敏度的图像采集,同时其配套的采集软件可以将DP30所采集的图像通过6种颜色的渲染,并合并成为一幅彩色图像。
主要参数 (DP30)
|
Items |
Specifications |
|
Camera type |
Single chip monochrome CCD camera with built-in shutter |
|
Imaging sensor |
2/3 inch monochrome CCD, 1.5 million (total) pixels, 1.45 million (effective) pixels, progressive scanning method (Sensor: ICX285AL, Pixel size: 6.45um x 6.45um) |
|
Cooling system |
Peltier cooling, CCD temperature: 5摄氏度 (constant) |
|
Camera mount |
C-mount |
|
CCD sensitivity |
Standard/NIR-M/NIR-H |
|
Gain |
x1, x2, x4, x8, x16 |
|
A/D |
12-bit |
|
Readout speed |
28MHz (live display with 1x1, 2x2, 4x4 binning), 14MHz (still image aquisition, live display with 8x8 binning) |
|
Full well capacity |
15,000e |
|
Readout noise |
8e |
|
Dark current |
0.75e/pixel/sec |
|
Quantum efficiency |
63% (550nm, Standard CCD sensitivity), 70% (600nm, NIR-H CCD sensitivity) |
|
Dynamic range |
1875:1 |
|
Exposure time |
600sec-0.1msec |
|
Background subtraction |
ON/OFF |
|
External trigger |
Event trigger input and trigger output (TTL compatible) |
|
Live Image frame rate |
1360x1024 (1x1), 680x512 (1x1), 680x510 (2x2), 340x250 (4x4), 170x120 (8x8) |
|
Still image acquisition time
(from start of shooting to display) |
1360x1024: approx. 0.3 sec, 680x510 (2x2): approx. 0.2 sec, 340x250 (4x4): approx. 0.1 sec, 170x120 (8x8): approx. 0.1 sec (minimum) |
|
Image display speed (frame rate) |
1360x1024: 15fps, 680x510 (2x2): 29fps 340x250 (4x4): 53fps, 170x120 (8x8): 53fps (maximum) |
|
Interface |
PCI bus interface, compliant with PCI Rev. 2.1 and 2.2 |
|
Operational conditions |
10度 – 35度, 20-80% (non condensing) |
|
Storage conditions |
-20度-60度, 10-90% |
配套软件主要性能 (DP-BSW)
|
Items |
Specifications |
|
Exposure control |
Exposure mode |
Auto, SFL auto, Manual |
|
AE lock |
Available |
|
Exposure adjustment |
Range: ?.0 EV, step: 1/3 EV |
|
Metering modes |
30%, 1%, 0.1% (metering area can be moved freely) |
|
Image integration |
Mode |
Integral/Average |
|
Number: |
64 frames (maximum) |
|
Black balance |
Auto (one-push, one-touch), manual |
|
Time-lapse photography |
Recording time: 1 sec. - 24hrs. 59min. 59sec., Number of images: 1000 Movie images can also be saved |
|
Movie recording |
Recording time: 1 sec. - 1hr. 59min. 59sec. |
|
Image file format |
TIFF (16/8bit), JPEG, BMP, PICT AVI, MPEG-1 |
|
Image orientation |
Flip/mirror/180?/TD> |
|
Contrast |
Linear, Hi, Mid, Lo |
|
Level adjustment |
Auto, manual |
|
Overlay display |
Scale, text, arrow, date |
|
Motorized microscope control |
Controls IX81/BX61 (Requires additional BX2-BSW or IX2-BSW)
· Motorized nosepiece
· Motorized mirror unit
· Excitation light shutter open/close |
|
OS |
Windows XP Professional Windows 2000 Professional |